HELGES – Varian & Spectro | Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES)
Contact:
Web:
Infrastructure belongs to:
The ICP-OES (Inductively Coupled Plasma Optical Emission Spectrometer) allows simultaneous detection and quantification of most elements in solutions at concentrations ranging from a few ng/g (ppb) to hundreds of µg/g (ppm). We analyse a wide selection of environmental samples for element concentrations, ranging from rock, soil, river water, seawater, and plant materials. We routinely analyse and quantify the following Elements: Li, Be, B, Na, Mg, Al, Si, P, S, K, Ca, Sc, Ti, Cr, Mn, Fe, Ni, Co, Cu, Zn, Ga, Ge, Rb, Sr, Ag, Cd, Ba, Pb, Bi.
The HELGES-laboratories are equipped with two types of optical spectrometers. The Varian 720-ES ICP-OES with an axial observation of the plasma offers high sensitivity and low detection limits, which is particularly suitable for low-concentration samples. The SpectroGreen ICP-OES has a radial plasma observation, which offers a high tolerance for samples with high salt loads and other complex matrices, thanks to the low matrix interferences.
Sample preparation For most applications we use ICP-OES analyses on solutions, hence, solids must be dissolved prior to analyses. There are several methods of dissolution using different acids (nitric, hydrochloric, hydrofluoric acid) or alkali fusion (e.g., NaOH). The amount of sample needed depends on the element of interest and if it is a major (wt%) or minor (ppm) component of the sample. Typically, for rocks or plants (wood, leaves) 50 – 100 mg powdered sample is needed, for water samples (stream or ground water) about 1 – 10 ml is used, which is typically further diluted to a minimum of 2-5 ml for quantitative simultaneous determination of cations. ICP-OES is a relative measurement method and therefore the instrument requires calibration prior to each measurement. To obtain accurate results the calibration standards need to be prepared in a similar matrix than the samples to be analysed. Hence, a rough estimate of the sample composition is required, which might be obtained by a first screening analyses by ICP-OES or by other analytical methods.
The capabilities of qualitative and quantitative detection varies strongly between different elements, and some elements cannot be analysed using ICP-OES. Levels of detection vary for each cation, ranging from 0.001 to 0.1 µg/ml.
Categories
Disciplinary Keywords
Instrumentation
Laboratory instrumentation
Instrument
-
Inductively Coupled Plasma Atomic Emission Spectrometer
Measurement method of atomic spectroscopy that uses an inductively-coupled plasma to excite and ionize atoms with measurement of the number and kind of ions in the plasma from their emission of electromagnetic radiation. (Source: IUPAC; https://iupac.org/wp-content/uploads/2019/10/PAC-REC-19-02-03.R2_PR191002MC.pdf)